NIST is currently working to establish a consortium to address challenges in the NVD program and develop improved tools and methods. You will temporarily see delays in analysis efforts during this transition. We apologize for the inconvenience and ask for your patience as we work to improve the NVD program.

CVE-2019-6540 Detail

Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.


Severity



CVSS 3.x Severity and Metrics:

NIST CVSS score
NIST: NVD
Base Score:  6.5 MEDIUM
Vector:  CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N


NVD Analysts use publicly available information to associate vector strings and CVSS scores. We also display any CVSS information provided within the CVE List from the CNA.

Note: NVD Analysts have published a CVSS score for this CVE based on publicly available information at the time of analysis. The CNA has not provided a score within the CVE List.

References to Advisories, Solutions, and Tools

By selecting these links, you will be leaving NIST webspace. We have provided these links to other web sites because they may have information that would be of interest to you. No inferences should be drawn on account of other sites being referenced, or not, from this page. There may be other web sites that are more appropriate for your purpose. NIST does not necessarily endorse the views expressed, or concur with the facts presented on these sites. Further, NIST does not endorse any commercial products that may be mentioned on these sites. Please address comments about this page to nvd@nist.gov.

Hyperlink Resource
http://www.securityfocus.com/bid/107544 Broken Link 
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01 Third Party Advisory  US Government Resource 

Weakness Enumeration

CWE-ID CWE Name Source
CWE-319 Cleartext Transmission of Sensitive Information cwe source acceptance level NIST   Reference acceptance level ICS-CERT  

Known Affected Software Configurations Switch to CPE 2.2

CPEs loading, please wait.

Denotes Vulnerable Software
Are we missing a CPE here? Please let us know.

Change History

4 change records found show changes

Quick Info

CVE Dictionary Entry:
CVE-2019-6540
NVD Published Date:
03/26/2019
NVD Last Modified:
11/03/2021
Source:
ICS-CERT